XY Stage Micrometers

Known lengths in two image directions.

Two-axis glass scales for checking horizontal and vertical image scale, axis alignment and magnification.

XY Stage Micrometers

Product selection

What it is and what it does

An XY stage micrometer places two defined scales at the specimen plane. It provides physical references for image scale in both directions and for checks of the relationship between the axes.

When to choose it

Choose it when both image directions must be checked under the same optical setup. Distributed grids are preferable when the task requires local scale across a wide field.

Check both axes at the specimen plane

Two-direction scale

Measure known spans along X and Y separately instead of assuming identical scale in both directions.

Axis relationship

Use the defined cross and axis geometry to check orientation. A short central scale does not establish accuracy across the whole field.

Calculate length per pixel or eyepiece division

Separate scale span, divisions and measurement records

Separate scale span, divisions and measurement records
ChoiceSelection basis
Scale spanChoose an observable span long enough for repeatable endpoint judgement within the usable field.
Division spacingFiner divisions are useful only if adjacent marks remain separable at the chosen objective and camera sampling.
Reported intervalsIdentify nominal dimensions, measured intervals and any uncertainty or traceability required for the result.

The stage scale is at the specimen plane. An eyepiece reticle is inside the viewing optics and requires a calibrated conversion to specimen length.

Choose spans and interpret measurement records

Where this product is used

Inspection and delivery

Inspect selected intervals along both axes, their intersection and orthogonality. Local interval error and accumulated span error are different quantities.

The supplied item is the selected patterned glass reference. The order identifies the measured intervals and accompanying measurement records.

Calibrate an eyepiece reticle with a stage scale

Selection notes

XY stage micrometers supply known graduations in two directions at the specimen plane.

Check each axis separately and record scale, axis direction and optical configuration. A single-axis check does not establish both scales.

Information for your inquiry

X/Y intervals, scale lengths, origin, orthogonality, format and calibration record.

Should Y be checked after X is calibrated?

Yes, when two-axis scale is needed. Separate checks can reveal directional differences introduced by sampling or configuration.