Custom Collimator & Autocollimator Reticles

Custom patterned glass reticles for collimators, visual autocollimators and optical alignment instruments, defined around the optical layout and mounting datums.

Instrument reticles

Reference patterns for the optical system

TL OPTIK manufactures drawing-defined glass reticles for projected references, visual receiving planes and replacement instrument assemblies. Pattern geometry, polarity and datum position are reviewed together so the finished component fits the intended optical and mechanical stack.

Choose this product when a crossline, ring, scale or custom reference must sit at a defined focal plane in a collimation or angle-measurement instrument. It is the patterned optical component, not a complete autocollimator, sensor or software system.

Custom glass reticle with crossline and concentric reference pattern for collimation instruments

Define the component role first

Define every reference plane

In a collimator, the patterned reference is placed near the objective focal plane and projected as a parallel beam. A visual autocollimator observes the reflected return image in a receiving plane; depending on the design, this can include an eyepiece reference reticle. Camera-based instruments may replace that receiving reticle with a CCD or CMOS sensor.

For this reason, “autocollimator reticle” is not one universal drawing. State whether the requested part is the projected pattern, the visual receiving reference or a replacement component copied from an existing instrument.

Pattern selection

Choose the pattern by the optical task

Pattern names alone are not enough. Line geometry, image sense and the relationship to the mating or return image must be defined for the assembled instrument.

Single crossline

A direct centre reference for projection, alignment and straightforward visual comparison.

Double crossline

A paired-line reference that can keep the return image visible between lines when matched to the instrument.

Concentric rings

Circular references for centring, symmetry checks and alignment around an optical axis.

Graduations and scales

Linear or angular references designed around focal length, magnification and the instrument reading convention.

Parallel lines and grids

Directional and two-axis references for alignment, image position and comparative inspection.

Apertures and custom patterns

Clear openings, targets, symbols or combined layouts manufactured from approved artwork.

Technical specifications

Drawing-defined configuration

These fields establish the specification structure. Final values and tolerances follow the approved drawing and agreed inspection method.

Instrument role
Collimator projection reticle, visual receiving reference, eyepiece graticule or drawing-defined replacement part
Pattern family
Single or double crossline, concentric rings, graduations, parallel lines, grid, aperture or custom artwork
Substrate and outline
Glass material, diameter or outline, thickness, edge finish and mounting features to the approved drawing
Clear aperture
Usable optical area and any edge exclusion defined separately from the outside diameter
Line geometry
Line width, clear spacing, double-line separation, scale interval and feature dimensions stated independently where critical
Pattern centre and datums
Pattern centre position relative to the outside diameter, mounting diameter, flats, holes or other mechanical datums
Orientation and image sense
Pattern rotation, viewing direction, film side and mirror-image requirement identified on the drawing
Polarity and illumination
Bright pattern on dark field or dark pattern on bright field, together with wavelength and illumination arrangement
Focal-plane relationship
Patterned surface and installed face located relative to the intended focal or image plane
Inspection and acceptance
Geometry, position, orientation, contrast and visible-defect limits follow the agreed drawing and inspection method

Optical-path fit

Match the full optical path

The patterned surface must be located and oriented relative to the focal plane. In autocollimation, a reflecting surface returns the projected reference to the receiving plane. A small mirror tilt produces a larger optical deflection; for a simple small-angle arrangement, x ≈ 2fθ describes the geometry between focal length, angle and return-image displacement.

This relation is not a complete accuracy specification. Instrument alignment, mechanics, sensor or observer and calibration remain part of the final measurement result.

Projected reticle
f
Reflecting surface
2θ
Receiving plane

Concept only — final optical layout and scale definition follow the instrument design.

Instrument applications

Where these reticles are used

Collimators and alignment telescopes

Projected crosslines, apertures and alignment references used to establish an optical axis or compare component position.

Visual autocollimators

Projected and receiving references for observing small angular changes from a reflected return image.

Optical test and metrology instruments

Custom references for checking tilt, alignment, squareness, parallelism or other system-level geometry.

TL OPTIK supplies the patterned glass reticle. The complete autocollimator, mechanical assembly, sensor, software and system calibration remain outside the component supply scope unless separately agreed.

Inquiry checklist

Send optical and mechanical context

A replacement sample helps identify the construction, but a controlled drawing is still needed to define what will be inspected and accepted.

  • Instrument type and whether the part is on the projection or receiving side
  • Pattern artwork, line geometry, polarity and current drawing revision
  • Outside dimensions, thickness, clear aperture and mounting datums
  • Pattern centre, rotation, film side, viewing direction and image sense
  • Focal-plane relationship, focal length and angular graduation convention where applicable
  • Illumination wavelength, bright-field or dark-field use and required contrast
  • Critical tolerances, inspection method, sample or mating-part information
  • Quantity, prototype approval route and delivery target

Separate component and instrument accuracy

Acceptance should separate pattern geometry from installed instrument performance. Inspect line width and spacing, pattern centre, orientation, polarity, clear aperture and visual defects against the approved drawing. Complete angle accuracy also depends on focal length, optical alignment, mechanics, detector or observer and calibration; it is not defined by reticle tolerance alone.

Common questions

Specify the reticle as part of the instrument

What is the difference between a collimator reticle and an autocollimator reticle?

A collimator reticle is placed near the focal plane and projected as a reference at infinity. An autocollimator uses a projected reference and observes its reflected return image. A visual instrument may also use an eyepiece or receiving reticle, while an electronic instrument may place a camera sensor at the receiving plane. The required component role must therefore be identified on the drawing.

How do I choose a single crossline or double crossline?

A single crossline provides a direct centre reference. A double crossline can create a clear comparison space for a mating or return image, but its line width and separation must suit the instrument magnification, illumination and expected image size.

Why must an angular graduation match the instrument focal length?

A linear displacement in the receiving focal plane represents an angle through the instrument focal length and measurement geometry. In a simple small-angle autocollimation arrangement, return-image displacement is approximately x ≈ 2fθ. The scale therefore cannot be transferred between instruments without reviewing focal length and the reading convention.

What information is needed for a replacement reticle?

Provide the existing component dimensions, mounting datums, patterned-side orientation, viewing direction, focal-plane location and clear images or measured artwork. A physical sample is useful, but the acceptance drawing should define the final geometry and orientation.

Does reticle tolerance define complete instrument angle accuracy?

No. Reticle geometry is one contributor. Complete instrument accuracy also depends on focal length, optical alignment, mechanical stability, detector or observer, signal processing and calibration. These should be verified at instrument level.