Centerline screens
Horizontal, vertical and crosshair references for centring and alignment. Define the intersection position and mounting direction.
Custom glass charts for optical comparators and profile projectors. Radius, angle, grid and profile patterns made to specified magnification and mounting dimensions.
TL OPTIK manufactures custom patterned glass charts for optical comparators and profile projectors. Used at the instrument’s viewing screen, these charts provide reference lines, radius arcs, angle graduations, grids or specified profiles for comparison with a projected workpiece image.
Chart geometry, markings and mounting dimensions are defined around the instrument configuration, intended projection magnification and inspection task. Chromium patterns are formed on glass by precision photolithography according to the approved drawing.


The instrument enlarges a workpiece edge or surface feature onto its viewing screen. The operator aligns the projected image with chart lines, arcs or profiles to compare dimensions, radii, angles and shape.
An optical comparator overlay is used over an existing screen. Patterned glass for a replacement screen assembly also needs the original screen-surface and mounting specifications. A transparent glass chart does not by itself provide a diffusing projection surface.
Use the chart name to identify the task, then define its actual geometry and scale. Toolroom charts can combine radius, angle and grid references on one glass plate.
Horizontal, vertical and crosshair references for centring and alignment. Define the intersection position and mounting direction.
Compare projected arcs with specified radii. Label the radius series and the magnification used to interpret it.
Protractor charts compare edge directions and included angles. Define the angular range, divisions and zero axis.
Compare diameters, concentricity and visible profile deviations. Formal roundness assessment needs an agreed measurement method.
X/Y references for positions and spacing. Specify the relationship between grid pitch and workpiece dimensions.
Compare the specified thread profile. Supply the thread form, pitch, projection direction and applicable standard.
Compare the specified tooth section using approved profile data and projection conditions.
Part outlines and agreed tolerance boundaries tied to the drawing revision, reference datums and magnification.
Pattern concepts only, not to scale. Thread and gear drawings illustrate chart types, not a specified standard profile.
Linear, diameter, radius and profile charts map workpiece dimensions onto the screen at a specified magnification. State whether chart labels show the workpiece value or the physical screen dimension.
At 10× magnification, a 1 mm workpiece feature projects to 10 mm on the screen.
1 mm × 10 = 10 mm
Illustrative geometry only. These examples are not a list of stock chart specifications or an instrument calibration.
Ideal uniform magnification preserves angles, but angular comparison still requires correct alignment and consideration of instrument distortion. A dimensional chart cannot be transferred between magnifications without reviewing its scale.
Choose line width for readable contrast and reliable edge comparison. Enlarging every feature of a small crosshair reticle by the same factor does not automatically produce a suitable screen chart.
Final values, tolerances and inspection methods follow the approved drawing. Confirm the viewing screen and chart as a matched assembly.
For optical comparator and profile projector manufacturers, metrology equipment suppliers and inspection laboratories. Send the information that controls the screen fit and comparison method.
Discuss your glass chartInspect agreed line widths, feature dimensions, scale spacing, centre position, orientation and visible defects against the approved drawing. Chart geometry inspection is separate from instrument calibration. Magnification, distortion, alignment, focus and the operator’s method contribute to the installed measurement result.
Quality, Inspection & DeliveryThis product is intended for the viewing-screen area of an optical comparator or profile projector. A measuring microscope reticle or eyepiece graticule uses a different mounting and optical arrangement. Specify the actual location of the component before ordering.
Dimensional, radius and profile patterns must match the intended projection magnification. Several labelled scales can share a chart, provided their correspondence is unambiguous. Angular references are unchanged by ideal uniform scaling, but alignment and instrument distortion still matter.
Not on diameter alone. An overlay sits over an existing viewing screen; a replacement screen may also require an image-forming or diffusing surface and a particular mounting construction. Review the original screen specification before defining a glass comparator screen.
A drawing, reference sample or clear image can support the review. Confirm dimensions, artwork, intended magnification, mounting references and acceptance criteria before manufacturing. Brand and model information alone does not establish compatibility.
Dimensional and pattern inspection records are defined with the order. If traceable calibration, a specified standard or a third-party certificate is required, agree the scope and feasibility before quotation. Component inspection does not certify the complete instrument.